APOGEE

Oxygen Sensor
Accuracy and Output


This sensor measures the amount of Oxygen (O2) gas
in air.
 
OXYGEN SENSOR

     Model SO-111:

$289

 
  OUTPUT, ACCURACY, ZERO OFFSET,
     & INFLUENCE FROM OTHER GASES
 
 
OUTPUT: 
The output signal of the Soil Sensor (-S) unit is approximately 50 mV in ambient air, or 2.4 mV per 1 % O2 and the output signal of the Fast-Response (-FR) unit is approximately 11 mV in ambient air, or 0.5 mV per 1% O2.

The user must calibrate their sensor at their location, at their pressure, to establish the exact multiplier to convert the mV reading to % O2. For example, if the output was exactly 11.0 mV the multiplier would be 1.905 (20.95 / 11.0).

Absolute accuracy of the sensor depends on how often it is calibrated. To achieve the best accuracy, automated calibration with ambient air at least once each day is recommended. Typical drift rates are less than 0.05% in 24 hours. 

ACCURACYAbsolute accuracy of the sensor depends on how often it is calibrated. To achieve the best accuracy, automated calibration with ambient air at least once each day is recommended. Typical drift rates are less than 0.05% in 24 hours.

ZERO OFFSET:  The mV output in ultra-pure nitrogen gas (0.000 % oxygen) is typically ± 0.05%. Precise measurements of hypoxic and anaerobic conditions can be made by making a periodic zero calibration of the sensor with ultra-pure nitrogen gas. The zero offset for each sensor is highly reproducible and should be entered into the software after a zero-test.

INFLUENCE FROM OTHER GASES:  The sensor is unaffected by CO, H2, and various acidic gases such as CO2, H2S, NOx, SOx. However, the Teflon membrane used in the sensor may be damaged by ozone.
 

INLINE OXYGEN SENSOR
      SPEC SHEETS
      PROGRAMS & MANUALS
      ORDER NOW


      UNDERSTANDING
        OXYGEN IN AIR

 

    OUTPUT, ACCURACY,
        ZERO OFFSET, & INFLUENCE
        FROM OTHER GASES
      THERMISTORS VERSUS
        THERMOCOUPLES