EVALUATION AND MODIFICATION OF COMMERCIAL INFRA-RED TRANSDUCERS FOR LEAF TEMPERATURE MEASUREMENTBruce Bugbee1, Matt Droter, Oscar Monje, and Bertrand Tanner²1
Crop Physiology Laboratory, Dept. of Plants
Soils, Biometeorology, Utah State University Logan, UT 84322-4820
|
BACK TO IRT PRODUCTS AND SPECIFICATIONS |
|
|
|
||
ABSTRACT |
TOP OF PAGE |
Accurate measurement of the leaf to air temperature gradient is crucial for the determination of stomatal conductance and other plant responses in both single leaves and in plant canopies. This gradient is often less than 1°C, which means that leaf temperature must be known to within about ±0.1°C. This is a challenging task, but new, miniature infra-red transducers from Exergen Corporation (Watertown, MA) and Everest Interscience (Tucson, AZ) can be modified and calibrated to achieve this accuracy. The sensors must be modified to add thermal mass and the Exergen sensor requires a measurement of sensor body temperature. Significant error is caused by the discharge of a capacitor in the standard Exergen sensor, but we tested it without the capacitor. The sensors respond rapidly to changes in target temperature, but require 2 to 10 minutes to respond to changes in sensor body temperature, which is often the largest source of error. A new, sensitive method for measuring field of view indicates substantial peripheral vision for both sensors and a wider field of view than specified by the manufacturers. Here we describe sensor output as a function of target and sensor body temperatures, and provide a generic (sensor independent) equation that can be used to achieve ±0.2 C accuracy with Exergen sensors. The equation was developed and verified using two black body calibrators. |
|
INTRODUCTION |
TOP OF PAGE |
Radiometric surface thermometers, more commonly known as infrared thermometers (IRTs), have many advantages over conventional thermometers for the measurement of surface temperatures, but they require consideration of target emissivity, field of view, and sensor body temperature. Equations for quantitatively determining the magnitude of these errors have been reviewed by Amiro et al. (1983). Conventional thermometers require a probe to be in physical contact with the leaf being measured, which is usually at a different temperature. Consequently, heat is transferred between the probe and the leaf until an equilibrium temperature is reached. The amount by which the probe is cooler or warmer than the leaf affects the temperature of the leaf and induces error in the measurement. In contrast, infrared thermometers do not have this intrusion error, but other sources of error must be corrected for to obtain true target temperature.Infrared thermometers are filtered to allow only a specific waveband, about 8 to 14 microns, to be transmitted to the IRT detector. This transmitted energy (E) is converted to temperature (T) via the Stefan-Boltzman Law which states E=esT4, where e is the emissivity of the object and s is the Stefan-Boltzmann constant (5.68 x 10-8 Joules m-2 s-1 K-4). Emissivity is defined as the radiation efficiency of a surface as compared to an ideal "black body" emitter. The power radiated by an object (emissivity) at thermal equilibrium in a vacuum must be equal to the power absorbed (absorbtivity). |
|
The following transducers were tested: |
Exergen IRt/c-271 |
Exergen IRt/c-0.21 |
Everest 3000AL2 |
|
POWER SUPPLY: |
none |
none |
± 5V |
OUTPUT: |
mV |
>mV |
mV |
COST: |
$ 199. |
$ 299. |
$ 795. |
SIZE: |
13 x 44 mm cylinder |
13 x 44 mm cylinder |
16 x 32 mm cylinder |
1Sensors from the Exergen Company (Watertown, MA) model IRt/c -K-27°C (wide angle) Four replicate sensors were tested model IRt/c .2-K-27°C (narrow angle) Four replicate sensors were tested. 2 Sensor from Everest Interscience, Inc. (Tucson, AZ; ph. 800-422-4342) model 3000 AL (narrow angle) |
|||
FIELD OF VIEW DETERMINATIONSensor FOV was determined by recording the target temperature as the IRT was pulled away from a black body cone. The opening of the black body had a diameter of 6 cm and was held at 60 °C. The perimeter of the cone was covered with low-emissivity aluminum foil, which, along with the surroundings, were at 20 °C. The sensor was first placed flush with the cone opening where it viewed 100% of the target and read 60°C. As the sensor was pulled away from the target it began to include the aluminum foil in its field of view. The ratio of target to perimeter at a given distance from the target was determined from the output of the transducer and the relative fractions of radiation emitted by the target and aluminum foil. A sensor with a wide FOV receives more energy from the perimeter at the same distance from the target than a narrow FOV sensor. The difference between the black body temperature and the observed IRT temperature was proportional to the percentage area of target and perimeter within the sensor FOV.BLACK BODY CALIBRATORAccurate calibration requires rigorous control of the sensor body
temperature in addition to control of the target black body temperature. A calibration
device that independently controlled sensor and target temperatures was built following
the design described by Kalma et al. (1988). The calibration unit consists of a separate
sensor block and a conical black body. The sensor block accommodates up to four sensors
simultaneously. The black body cone was 9 cm long x 3.8 cm diameter. The cone shape
increases the effective emissivity of the black body approximately by the ratio of the
surface area of the cone to the surface area of the opening (Kalma et al., 1988). The two
housings are separated thermally with 6 mm thick insulating material and nylon bolts. The
sensors were inserted into cylindrical holes in the sensor block facing the black body.
The temperature of the sensor bodies was measured by averaging thermocouples placed beside
each of the sensors inside the sensor holes. The temperature uniformity of the sensor
block was within ±0.02°C. Similarly, the black body temperature was measured by
averaging four thermocouples placed in 1 mm holes drilled in the top, sides, and bottom of
the conical housing.
|
RESULTS |
TOP OF PAGE |
SENSOR RESPONSE TIMEAll sensors responded quickly to changes in target
temperature, but slowly to changes in sensor body temperature (Table 1). Adding thermal
mass to the sensor body dampens rapid changes in temperature and thus improved accuracy in
fluctuating air temperature conditions. Response time to changing target temperature was
tested by repeatedly pointing the sensor at targets of different temperatures. Response
time to changing sensor body temperature was determined in the black body calibrator by
holding the target temperature constant and changing the sensor body temperature.
|
|
TABLE 1 |
TARGET |
SENSOR BODY |
EXERGEN WIDE |
< 1 s |
~ 120 s |
EXERGEN NARROW |
< 1 s |
~ 180 s |
EVEREST 3000AL |
< 1 s |
~ 600 s |
FIELD OF VIEWAll transducers had a wider field of view than advertised by the manufacturer, but there was no sharp distinction between target and nontarget radiation (Table 2, Figure 1). The fraction of the radiation from the edges may be acceptably small when the nontarget area is at a similar temperature to the target.FIGURE 1 The sensor field of view expressed as percent of the input signal coming from the target at increasing distances from the target. The target was a black body cone, which had low emissivity aluminum foil around its perimeter. All IRTs accurately measured the cone temperature when they were less than 0.2 diameters away from the target, but the fraction of the target viewed decreased as the distance from the target increased.TABLE 2 The field of view at which 90 or 99% of the signal comes from the target for the transducers in this study. The field of view as indicated by the manufacturer is shown in the 3rd column. Data from Figure 1. |
TABLE 2 |
90 % |
99 % |
MANUFACTURER |
EXERGEN WIDE |
101° | 138° | ~90° |
EXERGEN NARROW |
80° | 128° | ~35° |
EVEREST 3000AL |
51° | 115° | ~25° |
SENSOR BODY AND BLACK BODY TARGET ERRORSThe error in measurement of the black body target was within the manufacturers specifications for both transducers. The output of the Exergen was read as a thermocouple without additional linearization. The mV output of the Everest transducer was linearized with a 5th order polynomial supplied by the manufacturer.The Exergen sensor error was nonlinear at a constant sensor body temperature and changed from -0.5°C to 1.5 °C as the black body temperature was changed from -9 to +9°C above and below the sensor body temperature (Fig. 2). There was a slightly different curve at each sensor body temperature. The sensor error at all temperatures was less than 0.1°C when the transducer sensor body is held at the same temperature as the target.FIGURE 2 TOP GRAPH The errors in the measurement of black body temperature for 4 replicate Exergen IRt/c 0.2 sensors with a narrow field of view. The 4 sensors were purchased on different dates in 1995 and 1996. Each point represents the mean of 3 replicate readings taken on different days. Each data point was reproducible to within 0.02°C. The five groups of lines represent sensor body temperatures from 15 to 35°C. BOTTOM GRAPH Measurement error for a single Everest 3000 AL sensor. The four lines in the bottom graph represent sensor body temperatures from 20 to 35 °C. The mV output of the 3000 AL transducer was linearized with the fifth order polynomial supplied by the manufacturer. The error was reproducible on different days and could easily be corrected in software.LINEARIZATION OF THE EXERGEN NARROW FOV TRANSDUCERWe selected the Exergen narrow FOV sensor for further testing and
examined several types of equations to model the nonlinear relation between the sensor
error and black body temperature and found that the simplest relation was a parabolic
function (Eqn 2).
EXERGEN TRANSDUCER: RESIDUAL ERROR AFTER LINEARIZATIONFigure 4 shows the residual error after implementing the sensor
error correction term from Eqn 3 & 4 to the data in Figure 2.
|
DISCUSSION AND CONLCUSIONS |
TOP OF PAGE |
It was essential to measure and correct for the sensor body temperature of the Exergen IRTs. Increasing the sensor body mass dampened changes in sensor body temperature in fluctuating air temperatures and improved accuracy. A custom calibration equation for each sensor was needed to achieve ±0.1°C accuracy, but ±0.2°C accuracy was attained with a generic calibration equation.Consideration of the sensor FOV is necessary for optimal IRT placement, particularly with the wide FOV of the Exergen sensors. Our experience indicates that mounting the sensor vertically above the top of the canopy is the most desirable position for measuring canopy temperature. This position prevents the sensor from seeing sky temperature and the wide angle integrates a considerable canopy area.The Everest sensor came factory calibrated to ±1°C accuracy, and had a narrower FOV than the Exergen sensors, but would require the addition of considerable thermal mass to achieve ±0.2 °C accuracy in fluctuating air temperatures because of the slow response time of the sensor body.The Exergen sensors must be obtained without the capacitor to achieve the accuracy described in this paper.Some of the error is associated with the normal variation among different types of thermocouple wire. |
|
REFERENCES |
TOP OF PAGE |
Amiro, B. D., G. W. Thurtell, and T. J. Gillespie. 1983. A Small Infrared Thermometer for Measuring Leaf Temperature in Leaf Chambers. Jour. Of Experimental Botany. 34:1569-1576.Fuchs, M. 1990. Canopy Thermal Infrared Observations. Remote Sensing Reviews 5:323-333.Fuchs, M, and C.B. Tanner. 1966. Infrared thermometry of vegetation. Agron. Jour. 58: 597-601.Hatfield, J. L. 1990. Measuring Plant Stress with an Infrared Thermometer. HortScience, 25:1535-1538.Hipps, L. 1989. The Infrared Emissivities of Soil and Artemisia tridentata and Subsequent Temperature Corrections in a Shrub-Steppe Ecosystem. Remote Sens. Environ. 27:337-342.Huband NDS, and JL Monteith. 1986. Radiative surface temperature and energy balance of a wheat canopy. I. Comparison of radiative and aerodynamic temperature. Boundary-Layer Meteorology 36:1-17.Idso, S. B., RD Jackson, WL Ehrler, and ST Mitchell. 1969. A method for determination of infrared emittance of leaves. Ecology. 50:899-902.Kalma J. D., and H. Alksnis. 1988. Calibration of Small Infra-Red Surface Temperature Transducers. Ag. and Forest Meteorology, 43:83-98.Norman J. M., F. Becker. 1995. Terminology in thermal infrared remote sensing of natural surfaces. Ag. and Forest Meteorology. 77:153-166. |
|
APPENDIX A |
TOP OF PAGE |
12 GENERIC INSTRUCTIONS FOR MEASURING AND LINEARIZING AN EXERGEN IRt/c 0.2 TYPE K THERMOCOUPLE TRANSDUCER IN A CSI DATALOGGERBasic equation being implemented (see text): SEC = 0.25/Psb*[{(AppTarget T - Hsb)2}-Ksb]
|
|
Measure IRt/c apparent target temperature
|
Calculate correction factor
(SEC)
|